Optical metrology

Micro & Nanopositioners

Microscopy products

Version française

Amptek XR 100T CdTe

The XR-100T-CdTe represents a breakthrough in x-ray detector technology by providing "off-the-shelf" performance previously available only from expensive cryogenically cooled systems.

Features

  • CdTe Diode Detector
  • Thermoelectric (Peltier) Cooler
  • Cooled FET
  • Temperature Monitor
  • Beryllium Window
  • Hermetic Package (TO-8)

 

Applications

  • Medical X-Ray and Gamma Ray Detection: Mamography and Radiology
  • Characterization of x-ray tubes: direct output spectrum, current stability
  • X-Ray Fluorescence
  • Art and Archaeology
  • Portable X-Ray and Gamma Ray Instruments
  • X-Ray and Gamma Ray Research
  • Teaching
  • Nuclear Plant Monitoring
  • Uranium and Plutonium Detection

 

Model XR-100T-CdTe is a high performance x-ray and gamma ray detector, preamplifier, and cooler system using a 3 x 3 x 1 mm (or 5x5x1) Cadmium Telluride (CdTe) diode detector mounted on a two-stage thermoelectric cooler. Also on the cooler are the input FET feedback components to the charge sensitive preamplifier. The internal components are kept at approximately -30 ºC and can be monitored by an internal temperature sensor. The hermetic TO-8 package of the detector has a light tight, vacuum tight 4 mil (100 µm) Beryllium window.

All the critical connections between the detector and the preamplifier have been made internally to the XR-100T-CdTe to ensure quick, first time operation by the user. The XR-100T-CdTe is provided complete with BNC connectors and power cable.

 

Specifications

General
Detector type Cadmium Telluride (CdTe) Diode
Detector areas 3 x 3 mm (9 mm 2 ), 5 x 5 mm (25 mm 2 )
Detector thickness 1 mm
Energy resolution @ 122 keV,   57 Co 9 mm 2 : <1.2 keV FWHM, typical
25 mm 2 : <1.5 keV FWHM, typical
Dark counts <5 x 10 -3 counts/sec @ 10 keV < E < 1 MeV
Detector window Be, 4 mil thick (100 µm)
Preamplifier Charge Sensitive, with Current Divider Feedback
Case Size 3.00 x 1.75 x 1.13 in (7.6 x 4.4 x 2.9 cm)
Case weight 4.4 ounces (125 g)
Total power Less than 1 watt
Operation conditions 0°C to +40°C
Storage and Shipping Long term storage: 10+ years in dry environment
Typical Storage and Shipping: -20°C to +50°C, 10 to 90% humidity non condensing
Certification TUV Certification
Certificate #: CU 72072412 01
Tested to: UL 61010-1: 2004 R7 .05
CAN/CSA-C22.2 61010-1: 2004

Inputs
Preamp power ±8 volts @ 25 mA
Detector power +500 volts @ 1 µA
Cooler power
Current = 350 mA maximum
Voltage = 4 V maximum

Outputs
Preamplifier:
Sensitivity
Polarity

0.82 mV/keV
Negative signal out (1 kohm max. load)
Temperature monitor:
Sensitivity


PX5: direct reading in K through software.

 

Options

  • Other detector sizes (5x5x1mm) are available on special order
  • Other Be window thicknesses available on special order
  • Components for vacuum applications
  • Collimator Kit for high flux applications
  • Available in X-123CdTe configuration

 

 

Additional information

Amptek XR 100T CdTe PDF

 

If you would like more information, please contact us.

 

Home - Our products - Contact us - General conditions of sale